CAT III 1500V TRMS Clamp Meter
Performance Tests
13
K
iFlex Current
Probe (with
simulation)
1.05 mV @ 35 Hz NA 48.0 A 52.0 A
3 mV @ 100 Hz NA 48.0 A 52.0 A
15 mV @ 500 Hz NA 48.0 A 52.0 A
11.4 mV @ 20 Hz NA 921.0 A 979.0 A
57 mV @ 100 Hz NA 921.0 A 979.0 A
285 mV @ 500 Hz NA 921.0 A 979.0 A
28.8 mV @ 20 Hz NA 2323 A 2477 A
144 mV @ 100 Hz NA 2323 A 2477 A
720 mV @ 500 Hz NA 2323 A 2477 A
K
iFlex Current
Probe (with
52120A and
25-turn or
50-turn coil)
using
52120A/COIL3KA
using
52120A/COIL6KA
0.6 V @ 10 Hz 0.30 V @ 10 Hz 20 A 145.0 A 155.0 A
0.6 V @ 55 Hz 0.30 V @ 55 Hz 20 A 145.0 A 155.0 A
0.6 V @ 500 Hz 0.30 V @ 500 Hz 20 A 145.0 A 155.0 A
0.38 V @ 10 Hz 0.19 V @ 10 Hz 120 A 921.0 A 979.0 A
0.38 V @ 55 Hz 0.19 V @ 55 Hz 120 A 921.0 A 979.0 A
0.38 V @ 500 Hz 0.19 V @ 500 Hz 120 A 921.0 A 979.0 A
0.96 V @ 10 Hz 0.48 V @ 10 Hz 120 A 2323 A 2477 A
0.96 V @ 55 Hz 0.48 V @ 55 Hz 120 A 2323 A 2477 A
0.96 V @ 250 Hz 0.48 V @ 500 Hz 120 A 2323 A 2477 A
[1] When you apply a negative DC Voltage, the Product beeps and the backlight lights for 20 seconds. This is normal and you can
continue with the measurement.
[2] DC Voltage test points 700 V to 1450 V use 2x calibrator setup for high voltage or use DC power source 2290-5 rated
DC 5000 V.
[3] For 0 A to 20 A measurements, use a short from the calibrator AUX HI and LO for more accurate measurements.
[4] For 20 A to 120 A measurements, use a short from the 52120A HI CURRENT HI and LO for more accurate measurements.
Table 2. Performance Tests (cont.)
Test (Switch
Position)
Calibrator Output
52120A
Range
Readout Limit
Low High