Fluke 43B - General Specifications

Fluke 43B
54 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fluke 43B
Users Manual
46
Common Mode Voltage: 1000 V ac rms from earth ground, in compliance with
EN61010-1
Input Load Impedance (of host instrument): > 1M in parallel with up to 47
pF
Maximum Non-destructive Current: 1000 A
Duty Cycle: 0.5 A to 400 A continuous
Influence of Adjacent Conductor: < 9.0 mA/A
Influence of Conductor Position in Jaw Opening: ±1.0 % of reading +0.05 A
General Specifications
Output Cable Length: 2.5 m
Maximum Conductor Size: 32 mm
Storage Temperature: -20 °C to 60 °C
Operating Temperature: 0 °C to 50 °C
Relative Humidity: 10 °C to 30 °C: 95 %
30 °C to 40 °C: 75 %
40 °C to 50 °C: 45 %
Temperature Coefficient: 0.01 % x (specified accuracy)/ °C (< 18 °C or > 28
°C)
Altitude: Operating: 2000 m; 2000 m to 4000 m, derate category rating to
1000 V CAT II/600 V CAT III, Non-operating: 12000 m
Dimensions: 150 x 70 x 30 mm
Weight: 114 g