Calibrator
Menus
23
Remote Port Setup
Use the Remote Port Setup menu (inside the System Settings menu) to enable or disable
USBTMC, GPIB, Ethernet, and RS-232 ports by toggling a green/white switch. Tapping the
individual port buttons provides additional options and information. See the 5560A/5550A/5540A
Remote Programmers Manual at www.flukecal.com
.
The menu is:
•USB
This has USBTMC information:
• USB0::0x0F7E::0x800A::[serial number]::INSTR
Where:
• 0x0F7E: The Fluke vendor ID
• 0x800A: The Product ID number
• [serial number]: The serial number of the Product (the serial number is also on the rear
panel) or available using the *IDN? on one of the other remote interfaces. You can also
find it with the Setup > About menu.
• Restore USB Defaults with the Reset button.
• Ethernet
• DHCP (On or Off)
• Static IP Settings (IP Address, Gateway, Subnet Mask)
•MAC Address
•Port
• End of Line Character (CR/LF, CR, LF)
• Remote Interface (Terminal, Computer)
• Network Security Settings (Starting Address, Ending Address)
• Restore Ethernet Defaults
• GPIB
• GPIB Address
• Restore GPIB Defaults
• RS-232
• Data Bits (8, 7)
• Stop Bits (1, 2)
• Flow Control (None, RTS/CTS, XON/XOFF)
• Parity (None, Even, Odd)
• Baud Rate (9600, 19200, 38400, 57600, 115200)
• End of Line Character (CR/LF, CR, LF)
• Remote Interface (Terminal, Computer)
• Restore RS-232 Defaults