Verification Procedure
1. Set the DUT function to the position called for in Table 15.
2. Apply the indicated Calibrator output to the DUT for each test.
3. To verify the performance of the DUT, make sure that the DUT reads within the limits shown in the
Meter Response column.
Table 15. Performance Test - 77 Series IV DMM
Steps
Test
(Switch Position)
Calibrator Output
Meter Response
Lower Limit Upper Limit
1
e
Ohms
[2]
500 Ω 497.3 Ω 502.7 Ω
25 kΩ 4.974 kΩ 5.026 kΩ
3 50 kΩ 49.74 kΩ 50.26 kΩ
45 MΩ 4.974 MΩ 5.026 MΩ
5 10 MΩ 9.79 MΩ 10.21 MΩ
6 40 MΩ 39.19 MΩ 40.81 MΩ
7
R
Continuity
25 Ω Beeper On
8 250 Ω Beeper Off
9
K
AC Volts
50 mV 45 Hz 0.047 V ac 0.053 V ac
10 5 V 45 Hz 4.898 V ac 5.102 V ac
11 5 V 1 kHz 4.898 V ac 5.102 V ac
12 50 V 45 Hz 48.98 V ac 51.02 V ac
13 50 V 1 kHz 48.98 V ac 51.02 V ac
14 500 V 45 Hz 489.8 V ac 510.2 V ac
15 500 V 1 kHz 489.8 V ac 510.2 V ac
16 1000 V 45 Hz 978 V ac 1022 V ac
17 1000 V 1 kHz 978 V ac 1022 V ac
18
K
Hz
AC Volts Frequency
[1]
5 V 99 Hz 98.89 Hz 99.11 Hz
19 5 V 900 Hz 899 Hz 901 Hz
21 5 V 50 kHz 49.94 kHz 50.06 Hz
22
L
DC Volts
5 V 4.984 V dc 5.016 V dc
23 50 V 49.84 V dc 50.16 V dc
24 300 V 299.0 V dc 301.0 V dc
25 1000 V 996 V dc 1004 V dc
26 -1000 V -1004 V dc -996 V dc