Calibrator
Sample Applications
59
Table 16. Adjustment- 77 Series IV DMM
Function
(Switch Position)
Adjustment Step
[1]
Input Value
E
(DC Millivolts)
C-01 600.0 mV dc
C-02 120.0 mV dc
L
(DC Volts)
C-03 6.000 V dc
C-04 60.00 V dc
C-05 600.0 V dc
C
(AC Volts)
C-06 600.0 mV, 60 Hz
C-07 600.0 V, 60 Hz
e
(Ohms)
C-08 600.0 Ω
C-09 6.000 kΩ
C-10 60.00 kΩ
C-11 600.0 kΩ
C-12 6.000 MΩ
O
(Diode Test)
C-13 5.000 V dc
mA
(Milliamps)
C-14 400.0 mA dc
C-15 400.0 mA ac, 60 Hz
A
(Amps)
C-16 6.000 A dc
C-17 6.000 A ac, 60 Hz
[1] If the DMM is not connected correctly, or if the rotary switch is in the wrong position, the DMM beeps 2 times to alert the
user.