Users Manual
Testing Appliances
17
Using the Auto-Test Mode
The tester provides a number of factory programmed
auto-tests, see Table 3 (Class I appliances) and Table 4
(Class II appliances). An auto-test consists of a number of
single tests that will be carried out in the programmed
order. The test limits are pre-set, and the test result will
give a pass/fail indication.
See page 10 on how to create new auto-tests.
Auto-tests are locked out unless the Earth Bond test lead
resistance has been zeroed out, see page 7.
When any test fails during an auto-test, further tests can
not be carried out .
Performing Auto-Tests
You can run an auto-test in the standard mode or in the
fast mode. See ‘Test Modes: Standard or Fast’ on page
16 and ‘Selecting Fast or Standard Test Mode’ on page
10 for more information.
To start an auto-test, do the following:
1
AUTO
Select the auto-test mode and follow the
screen instructions.
The chapter ‘Description of the Tests’ on page 20
provides detailed information on the individual tests.
When the auto-test is finished a pass (
)
or a fail (
)
indication is displayed. Now you can review the results
before saving them, and save the results.