Process Calibrators
Specifications
17
Table 22. 717 Range and Resolution
1 15 30 100 300 500 1000 1500 3000 5000 10000
Range -1 to 1 -12 to
15
-12 to
30
-12 to
100
-12 to
300
0 to
500
0 to
1000
0 to
1500
0 to
3000
0 to
5000
0 to
10000
PSI 1.0000 15.000 30.000 100.00 300.00 500.00 1000.0 1500.0 3000.0 5000.0 10000
bar 0.0689 1.0342 2.0684 6.8947 20.684 34.474 68.947 103.42 206.84 344.74 689.48
mbar 68.948 1034.2 2068.4 6894.8 20684 34474 68948 NA NA NA NA
kPa 6.8948 103.42 206.84 689.48 2068.4 3447.4 6894.8 10342 20684 34474 68948
kg/cm2 0.0703 1.0546 2.1092 7.0307 21.092 35.153 70.307 105.46 210.92 351.53 703.06
cmH20@
4 °C
70.309 1054.6 2109.3 7030.9 21093 35154 70309 NA NA NA NA
cmH20@
20 °C
70.434 1056.5 2113.0 7043.4 21130 35217 70434 NA NA NA NA
inH20@
4 °C
27.681 415.12 830.42 2768.1 8304.2 13840 27681 41521 83042 NA NA
inH20@
20 °C
27.730 415.95 831.89 2773.0 8318.9 13865 27730 41595 83189 NA NA
mmHg@
0 °C
51.715 776.25 1551.5 5171.5 15515 25858 51715 NA NA NA NA
Overload Protection
Fuseless overvoltage protection
Temperature Coefficient
0.005 % of range per °C for temperature ranges of between −10 °C to 18 °C and 28 °C to
55 °C.
Loop Supply
24 V dc nominal
Pressure Module Input
Determined by pressure module