721
Users Manual
10
Home Menu Functionality
There are three options for the Home Menu:
P1/P2
mA/V/RTD
MENU
These options are shown across the bottom of the
display.
From anywhere within the menu structure, push to get
back to the Home Menu.
Backlight Use
Push to toggle on and off the backlight. This function
cannot be controlled by the serial interface.
Zero Function
In pressure mode and when the pressure is within the
zero limit, The Product zeros pressure on any port that is
currently shown on the display. The zero limits are within
10 % of the full-scale range of the selected sensor. If the
display shows “OL,” the zero function will not operate.
External Sensor (Absolute)
If the active display shows a selected absolute module,
and Z is pushed, push (SET) and you are prompted
to set or default the pressure reference. This is done with
and (the up and down arrows). The sensor port
must be open (vented) to atmosphere while this
procedure is done. Push (SET REF DONE) when
zero adjust is complete.

Other manuals for Fluke 721

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 721 and is the answer not in the manual?

Fluke 721 Specifications

General IconGeneral
Accuracy0.025 %
Built-in displayYes
Operating altitude0 - 2000 m
Connecting thread (male)1/8 \
Storage temperature (T-T)-20 - 60 °C
Pressure measurement range- bar
Operating temperature (T-T)-10 - 50 °C
Battery life35 h
Battery sizeAA
Battery technology-
Number of batteries supported4
Weight and Dimensions IconWeight and Dimensions
Depth58 mm
Width110 mm
Height200 mm
Weight539 g