Pressure Calibrator
Menus
21
RESOLUTION
To choose between a low or high resolution display:
1. Push to access the menus.
2. Push to step through the menu until
RESOLUTION is shown on the menu bar.
3. Push to select the Resolution menu.
4. Push or to turn low resolution on or off.
5. Push when finished.
hmq062.eps
HART
An internal 250 Ω HART resistor can be enabled when
the Product operates in the mA Measure-24 V mode. This
lets a HART Communicator be connected across the mA
terminals. It is not necessary to add an external resistor.
Note
When the HART resistor is on, the maximum
load-driving capacity is 750
Ω
.
1. Push to access the menus.
2. Push to step through the menu until HART is
shown on the menu bar.
3. Push to select the HART menu.
4. Push or to turn the HART Resistor on or off.
5. Push when finished.
hmq063.eps