8588A/8558A
Service Manual
10
Configuration: Z in 1 MΩ, 7.5 Digits, Internal Guard, Front Input, Aperture = 99 µs.
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
1000 V 0.000 0 V 0.019 0 V 0.019 0 V 0.002 4 V
1000 V 500.000 0 V 0.048 0 V 0.048 0 V 0.003 5 V
1000 V 1000.000 0 V 0.077 0 V 0.077 0 V 0.007 1 V
1000 V -500.000 0 V 0.048 0 V 0.048 0 V 0.003 5 V
1000 V -1000.000 0 V 0.077 0 V 0.077 0 V 0.007 1 V
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
100 V 0.000 00 V 0.001 90 V 0.001 90 V 0.000 23 V
100 V 100.000 00 V 0.007 70 V 0.007 70 V 0.000 70 V
100 V 200.000 00 V 0.013 50 V 0.013 50 V 0.001 40 V
100 V 0.000 00 V 0.001 90 V 0.001 90 V 0.000 23 V
100 V -100.000 00 V 0.007 70 V 0.007 70 V 0.000 70 V
100 V -200.000 00 V 0.013 50 V 0.013 50 V 0.001 40 V
1000 V 0.000 0 V 0.019 0 V 0.019 0 V 0.002 40 V
1000 V 500.000 0 V 0.048 0 V 0.048 0 V 0.003 50 V
1000 V 1000.000 0 V 0.077 0 V 0.077 0 V 0.007 10 V
1000 V -500.000 0 V 0.048 0 V 0.048 0 V 0.003 50 V
1000 V -1000.000 0 V 0.077 0 V 0.077 0 V 0.007 10 V