8588A/8558A
Service Manual
28
Configuration: Resolution 5 digits, LoI On
Optional Checks
DC Voltage (Rear Input)
Configuration: Z in Auto, 7.5 Digits, Internal Guard, Rear Input, Aperture = Auto
Suggested Reference: 5730A
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
1 mF 1. 000 00 mF 0.000 91 mF n/a 0.000 091 mF
1 mF 2.000 00 mF 0.001 70 mF n/a 0.000 220 mF
10 mF 2.000 0 mF 0.003 11 mF n/a 0.000 14 mF
10 mF 10.000 0 mF 0.010 38 mF n/a 0.000 83 mF
10 mF 20.000 0 mF 0.019 47 mF n/a 0.003 00 mF
100 mF 20.000 mF 0.031 08 mF n/a 0.001 4 mF
100 mF 100.000 mF 0.058 35 mF n/a 0.004 0 mF
100 mF 200.000 mF 0.103 80 mF n/a 0.010 0 mF
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
1 mF 0. 200 00 mF 0.000 29 mF n/a 0.000 014 mF
1 mF 1. 000 00 mF 0.000 91 mF n/a 0.000 068 mF
1 mF 2.000 00 mF 0.001 70 mF n/a 0.000 140 mF
10 mF 2.000 0 mF 0.003 11 mF n/a 0.000 14 mF
10 mF 10.000 0 mF 0.010 38 mF n/a 0.000 68 mF
10 mF 20.000 0 mF 0.019 47 mF n/a 0.001 40 mF
100 mF 20.000 mF 0.031 08 mF n/a 0.001 5 mF
100 mF 100.000 mF 0.058 35 mF n/a 0.004 5 mF
100 mF 200.000 mF 0.103 80 mF n/a 0.018 0 mF
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
10 V 0.000 000 V 0.000 001 V 0.000 001 V 0.000 000 30 V
10 V 10.000 000 V 0.000 037 V 0.000 054 V 0.000 006 40 V