Reference Multimeter / 8½ Digit Multimeter
Calibration Verification
23
Resistance Normal
Configuration: 7.5 Digits, Aperture = Auto, 4-wire (2-wire at 100 MΩ), Lo Current Off, Internal Guard,
Front Input, Filter Off, Zeros ≤10 MΩ performed into a 4-wire short circuit; Zeros >10 MΩ performed
across the resistor with the current shorted as described in the Operators Manual.
Suggested Reference: 5730A
Range Frequency Reference Value
Permissible Error (±)
Fluke
Uncertainty (±)
8588A 8558A
1 A 30 kHz 1.000 000 A 0.000 404 A 0.000 507 A 0.000 028 A
1 A 1 kHz 2.000 000 A 0.000 799 A 0.000 903 A 0.000 110 A
30 A 10 Hz 10.000 000 A 0.008 50 A n/a 0.000 72 A
30 A 55 Hz 10.000 000 A 0.010 95 A n/a 0.000 68 A
30 A 305 Hz 10.000 000 A 0.010 95 A n/a 0.000 50 A
30 A 1 kHz 10.000 000 A 0.010 95 A n/a 0.000 50 A
30 A 2 kHz 10.000 000 A 0.010 95 A n/a 0.000 50 A
30 A 5 kHz 10.000 000 A 0.010 95 A n/a 0.000 50 A
30 A 10 kHz 10.000 000 A 0.010 95 A n/a 0.000 50 A
30 A 1 kHz 30.000 000 A 0.046 38 A n/a 0.002 30 A
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
1 Ω 0.000 000 0 Ω 0.000 005 2 Ω 0.000 005 8 Ω 0.000 000 10 Ω
1 Ω 1.000 000 0 Ω 0.000 019 2 Ω 0.000 025 8 Ω 0.000 002 40 Ω
1 Ω 1.900 000 0 Ω 0.000 031 8 Ω 0.000 043 8 Ω 0.000 004 90 Ω
10 Ω 0.000 000 Ω 0.000 018 Ω 0.000 026 Ω 0.000 001 0 Ω
10 Ω 10.000 000 Ω 0.000 117 Ω 0.000 186 Ω 0.000 027 0 Ω
10 Ω 19.000 000 Ω 0.000 031 Ω 0.000 330 Ω 0.000 034 0 Ω
100 Ω 0.000 00 Ω
0.000 06 Ω 0.000 07 Ω 0.000 010 Ω
100 Ω 100.000 00 Ω 0.000 98 Ω 0.001 37 Ω 0.000 150 Ω
100 Ω 190.000 00 Ω 0.001 81 Ω 0.002 54 Ω 0.000 270 Ω
1 kΩ 0.000 000 0 kΩ 0.000 00 6 kΩ 0.000 000 7 kΩ 0.000 000 10 kΩ
1 kΩ 1.000 000 0 kΩ 0.000 009 7 kΩ 0.000 013 7 kΩ 0.000 001 20 kΩ
1 kΩ 1.900 000 0 kΩ 0.000 017 9 kΩ 0.000 025 4 kΩ 0.000 001 90 kΩ
10 kΩ 0.000 000 kΩ 0.000 006 kΩ 0.000 007 kΩ 0.000 001 0 kΩ
10 kΩ 10.000 000 kΩ 0.000 098 kΩ 0.000 137 kΩ 0.000 015 0 kΩ
10 kΩ 19.000 000 kΩ 0.000 181 kΩ 0.000 254 kΩ 0.000 027 0 kΩ