Reference Multimeter / 8½ Digit Multimeter
Calibration Verification
15
Digitize DC Current
Configuration: Aperture = 200 ns, Trigger count 101010, Low pass filter Off.
Configuration: Aperture = 200 ns, Trigger count 101010, Low pass filter 100 kHz.
Configuration: Aperture = 200 ns, Trigger count 101010, Low pass filter 3 MHz.
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
10 mA 0.000 000 00mA 0.001 600 mA 0.001 600 mA 0.000 280 mA
10 mA 10.000 000 mA 0.002 620 mA 0.002 620 mA 0.000 280 mA
10 mA -0.000 000 mA 0.001 600 mA 0.001 600 mA 0.000 280 mA
10 mA -10.000 000 mA 0.002 620 mA 0.002 620 mA 0.000 280 mA
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
10 mA 0.000 000 0 mA 0.000 900 mA 0.000 900 mA 0.000 230 mA
10 mA 1.000 000 0 mA 0.001 520 mA 0.001 520 mA 0.000 240 mA
10 mA -0.000 000 0 mA 0.000 900 mA 0.000 900 mA 0.000 230 mA
10 mA -1.000 000 0 mA 0.001 520 mA 0.001 520 mA 0.000 240 mA
Range Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
10 mA 0.000 000 0 mA 0.000 900 mA 0.000 900 mA 0.000 230 mA
10 mA 1.000 000 0 mA 0.001 520 mA 0.001 520 mA 0.000 240 mA
10 mA 0.000 000 0 mA 0.000 900 mA 0.000 900 mA 0.000 230 mA
10 mA -1.000 000 0 mA 0.001 520 mA 0.001 520 mA 0.000 240 mA