Reference Multimeter / 8½ Digit Multimeter
Calibration Verification
21
Configuration: 7.5 Digits, Internal Guard, Front Input, DC Coupled, Wideband. RMS Filter 40Hz.
Suggested Reference: 5730A
Range Frequency Reference Value
Permissible Error (±)
Fluke Calibration
Uncertainty (±)
8588A 8558A
1 kV (AC) 1 kHz 1000.000 V 0.310 V 0.310 V 0.021 0 V
1 kV 2 kHz 1000.000 V 5.100 V 5.100 V 0.021 0 V
1 kV 10 kHz 1000.000 V 19.400 V 25.900 V 0.021 0 V
Range Frequency Reference Value
Permissible Error (±)
Fluke
Uncertainty (±)
8588A 8558A
10 μA 10 Hz 10.000 00 μA 0.029 13 μA 0.036 25 μA 0.001 80 μA
10 μA 55 Hz 10.000 00 μA 0.029 13 μA 0.036 25 μA 0.002 10 μA
10 μA 305 Hz 10.000 00 μA 0.029 13 μA 0.036 25 μA 0.001 80 μA
10 μA 1 kHz 10.000 00 μA 0.029 13 μA 0.036 25 μA 0.001 80 μA
10 μA 2 kHz 10.000 00 μA 0.029 13 μA 0.036 25 μA 0.001 80 μA
10 μA 5 kHz 10.000 00 μA 0.029 13 μA 0.036 25 μA 0.001 80 μA
10 μA 10 kHz 10.000 00 μA 0.029 13 μA 0.036 25 μA 0.001 80 μA
10 μA 1 kHz 20.000 00 μ
A 0.055 03 μA 0.068 63 μA 0.003 50 μA
100 μA 10 Hz 100.000 0 μA 0.040 0 μA 0.052 9 μA 0.008 0 μA
100 μA 55 Hz 100.000 0 μA 0.040 0 μA 0.052 9 μA 0.007 4 μA
100 μA 305 Hz 100.000 0 μA 0.040 0 μA 0.052 9 μA 0.006 6 μA
100 μA 1 kHz 100.000 0 μA 0.040 0 μA 0.052 9 μA 0.006 6 μA
100 μA 2 kHz 100.000 0 μA 0.040 0 μA 0.052 9 μA 0.006 6 μA
100 μA 5 kHz 100.000 0 μA 0.072 3 μA 0.091 6 μA 0.006 8 μA
100 μA 10 kHz 100.000 0 μA 0.072 3 μA 0.091 6 μA 0.008 1 μA
100 μA 20 kHz 100.000 0 μA 0.099 4 μA 0.118 7 μA 0.015 0 μA
100 μA 30 kHz 100.000 0 μA 0.099 4 μ
A 0.118 7 μA 0.027 0 μA
100 μA 1 kHz 200.000 0 μA0.073 5 μA 0.092 9 μA 0.009 6 μA
1 mA 10 Hz 1.000 000 mA 0.000 400 mA 0.000 529 mA 0.000 063 mA
1 mA 55 Hz 1.000 000 mA 0.000 400 mA 0.000 529 mA 0.000 055 mA
1 mA 305 Hz 1.000 000 mA 0.000 400 mA 0.000 529 mA 0.000 055 mA
1 mA 1 kHz 1.000 000 mA 0.000 400 mA 0.000 529 mA 0.000 055 mA
1 mA 2 kHz 1.000 000 mA 0.000 400 mA 0.000 529 mA 0.000 055 mA
1 mA 5 kHz 1.000 000 mA 0.000 723 mA 0.000 916 mA 0.000 055 mA