Reference Multimeter / 8½ Digit Multimeter
Calibration Interval and Performance
5
10. Continue through the adjustment process until all steps are complete.
The last adjustment for 8558A is: Step 703: Current Zero. The last step for 8588A is Step 704:
Capacitance zero.
Table 2. Table Points
Step Description Input
201 ADC Reference Characterize N/A
202 10 mV Offset 0 mV
203 10 mV Gain +10 mV
204 100 mV Offset 0 mV
205 100 mV Gain +100 mV
206 1 V Offset 0 V
207 1 V Gain +1 V
208 10 V Offset 0 V
209 10 V Gain +10 V
210 100 V Offset 0 V
211 100 V Gain +100 V
212 1 kV Offset 0 V
213 1 kV Gain +1 kV
214
ACV 1 MΩ HF 1 kHz Ref 100 V 1 kHz
215
ACV 1 MΩ HF 60 kHz 100 V 60 kHz
301 100 μA Offset 0 μA
302 100 μA Gain +100 μA
303 1 mA Offset 0 mA
304 1 mA Gain +1 mA
305 10 mA Offset 0 mA
306 10 mA Gain +10 mA
307 100 mA Offset 0 mA
308 100 mA Gain +100 mA
309 1 A Offset 0 A
310 1 A Gain +1 A
311
[1]
10 A Offset 0 A
312
[1]
10 A Gain +10 A