Introduction and Specifications
Specifications
1
1-11
Conductance [±(percent of reading + counts)]
Range Resolution Accuracy
Open Circuit
Voltage
300.00 nS 0.01 nS 0.5% + 20 3.2V
3000.0 nS 0.1 nS 0.5% + 20 3.2V
Overload Protection: 1000V rms
Capacitance [±(percent of reading + counts)]
Range Resolution Accuracy
10000 pF
1
10 pF 1.9% + 20
.1000 µF 100 pF 1.9% + 2
1.000 µF 1 nF 1.9% + 2
10.00 µF 0.01 µF1.9% + 2
100.0 µF0.1 µF1.9% + 2
1000 µF1 µF1.9% + 2
10000 µF
2, 3
100 µF 10% + 900
Overload Protection: 1000V rms
1. 10,000 pF range last digit reads zero.
2. 10,000 µF range last two digits read zero.
3. Using REL to zero internal offset.
Continuity Beeper Values
Range Beeper On Beeper Off
300 32 136
3 k 212 725
30 k 1586 4799
300 k 15.3 k 45.5 k
3 M 152.7 k 459.1 k
30 M 66 k 194 k
Resistance [±(percent of reading + counts)]
Range Resolution Accuracy Maximum
Current
Open Circuit
Voltage
300.00 0.01 0.07% + 10 1 mA 3.2V
3.0000 k 0.1 0.07% + 2 120 µA1.5V
30.000 k 1 0.07% + 2 14 µA1.5V
300.00 k 10 0.07% + 2 1.5 µA1.5V
3.0000 M 100 0.15% + 2 150 nA 1.5V
30.000 M 1 k 0.2% + 3 320 nA 3.2V