Service Manual
4-20
Rel Test/Touch Hold Test 4-17.
1. Turn on the GMM to VDC, and apply 1000V dc.
2. Press t[Rel].
3. Check for a Rel reading between the limits shown in Table 4-19.
Table 4-19. Rel Performance Test
Input Input Source Lower Limit Upper Limit
1000.0V dc 5700A using GMM volts input -0.2 0.2
4. Press é.
5. On the 5700A, press l to select Standby.
6. Check for a GMM reading within the limits shown in Table 4-20.
Table 4-20. Touch Hold Performance Test
Input Input Source Lower Limit Upper Limit
0V dc (open) -0.2 0.2