Fluke 867B - DC Volts Performance Test; DC Volts Test

Fluke 867B
101 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
867B,863
Service Manual
4-6
DC Volts Test 4-6.
Connect the GMM and 5700A in Configuration 2 (Figure 4-2). Then perform the three
steps called for in Table 4-3. Before proceeding to the next test, place the 5700A in
Standby.
Table 4-3. DC Volts Performance Test
Model 867B Model 863
Step
GMM
Range
Input Input Source
Lower
Limit
Upper
Limit
Lower
Limit
Upper
Limit
1 30.000V 29.000V 5700A 28.991 29.009 28.986 29.014
2 300.00V -290.00V 5700A -290.09 -289.91 -290.14 -289.86
3 1000.0V 1000.0V 5700A 999.5 1000.4 999.4 1000.6
HI