Maintenance
Replacing the 440 mA Fuse
3
3-9
Replacing the 440 mA Fuse 3-7.
Refer to Figure 3-3 when replacing the 440 mA fuse.
Spare Fuse
Fuse
Spare Fuse
os7f.eps
Figure 3-3. Replacing the 400 mA Fuse
Replacing the 11A (High Energy) Fuse 3-8.
Follow the “Disassembly” instructions in this chapter (Figure 3-1). Remove this fuse
(shown as step 9 in Figure 3-3.) Check for signs of damage (smoked, arced areas) on the
circuit board.
Note
Avoid switching 440 mA and 11A fuses. The 440 mA fuse is accessible
through the battery compartment and does not require GMM disassembly.
The 11A fuse resides on the opposite side of the GMM circuit and does
require disassembly for access.