Perfomance Testing and Calibration
Performance Tests
4
4-11
HI
HI
LO
LO
HI
OUTPUT
V A
SENSE
V
AUX
CURRENT
GUARD GROUND
WIDEBAND
5700A Calibrator
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 53 4
ACOMVEXT TRIG
mA/
A
os16f.eps
Figure 4-4. Configuration 5 (Ohms)
Capacitance Test 4-10.
Rotate the GMM selector to Diode/Capacitance Test (
ä). Connect the GMM Ë
and â inputs to the GenRad 1412BC. Then perform the test steps in Table 4-9.
Table 4-9. Capacitance Performance Test
Step GMM Range Input Input Source Lower Limit Upper Limit
1 10000 pF open open 0 130
21 µF 0.950 µF GenRad 1412-BC 0.930 0.970