Perfomance Testing and Calibration
Calibration
4
4-21
Glitch Capture Test 4-18.
1. Turn on the GMM to mV.
2. Press , vc[View], w [Glitch Capture], t [Arm].
3. Apply a .070V p-p Pulse 10 kHz Pulse wave.
4. Check for the Glitch Capture PASS/FAIL indications shown in Table 4-21.
Table 4-21. Glitch Capture Performance Test
Input Input Source PASS FAIL
.07V p-p Pulse, 10 kHz, 50% duty PM5139 to the GMM volts input Trigger occurs No Trigger
External Trigger Test 4-19.
1. Turn on the GMM to mV DC.
2. Press: , v [View], u [Trigger], w [Internal/External](to select
external), x [Exit].
3. Press v [Single Shot] , t [Arm].
4. Apply a 2V p-p 100 kHz Pulse wave.
5. Check for the PASS/FAIL indications shown in Table 4-22.
Table 4-22. External Trigger Performance Test
Input Input Source PASS FAIL
2.0V p-p, 100 kHz Pulse, 50% duty PM5139 to the GMM EXT TRIG input. Trigger occurs No Trigger