Service Manual
4-10
Ohms and Conductance Tests 4-9.
Rotate the GMM selector to Ohms (
ò). Connect the GMM and the 5700A in
Configuration 5 (Figure 4-4). Then perform the test steps in Table 4-7 (for ohms) and
Table 4-8 (for conductance).
Before proceeding to the next test, place the 5700A in Standby.
Table 4-7. Ohms Performance Test
Step Range Input Input Source Lower Limit Upper Limit
1 300.00 0.00 Shorting Bar 0.00 0.10
2 300.00 190.00 5700A 189.77 190.23
3 3.0000 k 1.9000 k 5700A 1.8985 1.9015
4 30.000 k 19.000 k 5700A 18.985 19.015
5 300.00 k 190.00 k 5700A 189.85 190.15
6 3.0000 M 1.9000 M 5700A 1.8970 1.9031
7 30.000 M 10.000 M 5700A 9.977 10.023
Table 4-8. Conductance Performance Test
Step Range Input Input Source Lower Limit Upper Limit
1 300.00 nS 10.00 M 5700A 99.30 100.70
2 3000.0 nS 1.0000 M 5700A 993.0 1007.0