Perfomance Testing and Calibration
Performance Tests
4
4-17
Table 4-16. AC Amps Average Performance Test
Step
GMM Range
Input
Input Source Lower Limit Upper Limit
1** 300.0 µA
(863 na)
short 0.0 0.4
2 300.0 µA
(863 na)
210.0 µA, 10 kHz 5700A 205.4 214.6
3 3.000A 2.90A, 3 kHz * 5725A 2.874 2.926
* Use 2.90A, 1 kHz with Fluke 5500A.
** When using the battery eliminator the upper limit for this test is 2.0.