Perfomance Testing and Calibration
Performance Tests
4
4-7
Diode Test 4-7.
Rotate the GMM selector to Diode Test (
ä). Connect the GMM and 5700A in
Configuration 2 (Figure 4-2,
äselected). Then perform the three steps called for in
Table 4-4.
Before proceeding to the next test, place the 5700A in Standby.
Table 4-4. Diode Test Performance Test
Step
GMM Range
Input Input Source
Lower Limit Upper Limit
1 Auto 2.9V 5700A 2.3 3.5
2 -2.9V 5700A -2.3 -3.5
3 Manual 2.5000V 5700A 2.4986 2.5015