Perfomance Testing and Calibration
Performance Tests
4
4-15
Logic Test (867B Only) 4-13.
Rotate the GMM selector to LOGIC. If necessary, press t so that ‘TTL’ is selected
(highlighted). Connect the GMM and the 5139 in Configuration 9 (Figure 4-7.) Then
perform the test shown in Table 4-13.
Table 4-13. Logic Performance Test
Range Input Input Source Lower Limit Upper Limit
na 2.0 MHz, 1.5V p-p pulse,
50% duty cycle, 0.7V dc
offset*
PM 5139 1.9989 2.0011
* 1.5V p-p transitioning from 0.7V to 2.2V.
FREEZE
DISPLAY
MODE
SAVE
PRINT
MIN
MAX
RANGE
WAKE UP
HZ
TOUCH
HOLD
12 53 4
ACOM
V
EXT TRIG
mA/
A
LOCAL
ADDR
x10
OUTPUT
Zo 50
PM
5139 function generator 0.1 mHz 10MHz
RESET
RANGE
A
mV
V
V
Hi
Z
mA
A
LOGIC
os31f.eps
Figure 4-7. Configuration 9 (PM5139)