7-10 Voltage Reference
:CALibration :VOLTage :LOAD :FACTory
Restore Factory Calibration
Recalls the calibration factors stored by the factory before delivery or after service.
:CALibration :VOLTage :LOAD :USER
Restore User Calibration
Recalls the calibration factors last stored by the user.
:CALibration :VOLTage :STORe
8<YYYYMMDD>
Store User Calibration
Ends the calibration process and stores the calibration factors using the date code YYYYMMDD, that is year (4
digits), month (2 digits), and day (2 digits).
Example:
SEND® :CAL:VOLT:STOR 20040731