EasyManua.ls Logo

Fluke PM6690 - Rear Inputs;Outputs; Measuring Functions

Fluke PM6690
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Trig level = 0.5V
DC coupling
The standard deviation (
Std) should be <100 ps.
Rear Inputs/Outputs
10 MHz OUT
Connect an oscilloscope to the 10 MHz output on the
rear of the counter. Use a coaxial cable and 50 W termi
-
nation.
The output voltage should be sinusoidal and above 1
V
RMS
(2.8 V
p-p
).
EXT REF FREQ INPUT
Recall the DEFAULT settings.
Connect a stable 10 MHz signal (e.g REF OUT from an
-
other counter) to input A.
Connect a 10 MHz, 100 mV
RMS
, (0.28 V
P-P
) signal from
the LF synthesizer to EXT REF IN.
Select Ext Ref. by keying in the following sequence:
SETTINGS ® Timebase Ref ® Ext
The display should show 10 MHz.
Change the external reference frequency to 5 and
1 MHz.
The counting should continue, and the display should
still show 10 MHz.
EXT ARM INPUT
Proceed from the test above.
Select
MANual trigger.
Settings for the pulse generator:
single shot pulse, manual trigger,
amplitude TTL = 0 - 2 V
PP
, and
duration = 10 ns.
Connect the pulse generator to Ext Arm Input.
Activate start arming by keying in the following se
-
quence:
SETTINGS ® Start Chan ® E
The counter does not measure.
Apply one single pulse to Ext Arm Input.
The counter measures once and shows 10 MHz on the
display.
Measuring Functions
Connect a 10 MHz sine wave signal with approx.
1V
RMS
amplitude into 50 W via a power splitter to Input
A and Input B, e.g. from 10 MHz Out on the rear panel.
Recall the DEFAULT settings.
Select the following settings for the timer/count er via
INPUT
A
and INPUT B:
50 W impedance for A and B.
MANual trigger.
2-6 Performance Check, Rear Inputs/Outputs
Selected Function Action Display P/F
FREQ A
10 MHz
2)
FREQ B
10 MHz
2)
FREQ C
--------------
3)
FREQ RATIO A/B
1.0000000
FREQ RATIO C/B
0.0000000
3)
PER SINGLE A
100 ns
2)
PER SINGLE B
100 ns
2)
PER AVERAGE A
100 ns
2)
PER AVERAGE B
100 ns
2)
PULSE POS A
50.000 ns
1)
PULSE NEG A
50.000 ns
1)
TIME INT A to B
0±1ns
Select NEG
SLOPE B
50.000 ns
1)
RISE TIME A
Select AUTO
trigger
30.000 ns
2)
FALL TIME A
30.000 ns
2)
PHASE A rel B
180° or -180°
1)
PHASE B rel A
180° or -180° 1)
PHASE A rel A
DUTY POS A
0.500000
1)
DUTY NEG A
0.500000
1)
VOLT MAX A
+0.75 V
2)
VOLT MIN A
-0.75 V
2)
Table 2-8 Measuring functions check
1) Value depends on the symmetry of the signal.
2) Exact value depends on input signal.
3) If an RF option is installed.

Table of Contents

Other manuals for Fluke PM6690

Related product manuals