EasyManua.ls Logo

Fluke PM6690 - Trigger Indicators and Input Controls; Reference Oscillators and Acceptance Test

Fluke PM6690
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Press
INPUT A and select MANual trigger level and 50 W
input impedance.
Connect the LF synthesizer to Input A. Use the follow
-
ing settings (into 50 W):
Sine, 10 kHz, 0.9 V
PP
, and +0.50 V
DC
.
Verify that the three modes for the trigger indicator are
working properly by changing the trigger level:
Press the
Trig key and enter +1 V via the keyboard,
then verify by pressing
EXIT/OK. Check the trigger
indicator according to Table 2-4.
Press the
Trig key and enter –1 V via the keyboard
by pressing the
± key, then verify by pressing
EXIT/OK. Check the trigger indicator according to
Table 2-4.
Press the
Trig key and enter 0 via the keyboard, then
verify by pressing
EXIT/OK. Check the trigger indica
-
tor according to Table 2-4.
Apply the signal to Input B instead.
Press
MEAS FUNC ® Freq ® Freq (A) ® B
Press
INPUT B and select MANual trigger level and
50 W input impedance..
Repeat the trigger level settings above to verify the three
trigger indicator modes for Input B.
Input Controls
Recall the DEFAULT settings.
Connect the LF synthesizer to Input A. Use the same
settings as in the previous test.
Press
INPUT A and select DC and 50 W.
Press
EXIT/OK.
Check the V
max
and V
min
voltage levels on the display
according to the first row in Table 2-6.
Perform the rest of the settings in sequence, and read the
corresponding V
max
and V
min
values. Remember that all
these values are approximate and serve only as indica
-
tors of state changes.
Connect the generator to Input B.
Press
MEAS FUNC ® Freq ® Freq (A) ® B
Press
INPUT B and select DC and 50 W.
Press
EXIT/OK.
Check the V
max
and V
min
voltage levels on the display
according to the first row in Table 2-6.
Perform the rest of the settings in sequence, and read the
corresponding V
max
and V
min
values.
Reference Oscillators
X-tal oscillators are affected by a number of external condi
-
tions like ambient temperature and supply voltage. Aging is
also an important factor. Therefore it is hard to give limits for
the allowed frequency deviation. The user himself must de
-
cide the limits depending on his application, and recalibrate
the oscillator accordingly.
To check the accuracy of the oscillator you must have a cali
-
brated reference signal that is at least five times more stable
than the oscillator that you are testing. See Table 2-7 and the
list of test equipment on page 2-2. If you use a non-10 MHz
reference, you can use the mathemati cs in the timer/counter to
multiply the reading.
Recall the DEFAULT settings. See page 2-2
Connect the reference to input A
Check the readout against the accuracy requirements of
your application.
n
Acceptance Test
Table 2-7 can serve as an acceptance test and gives a worst
case figure after 30 minutes warm-up time. All deviations that
can occur in a year are added together.
Resolution Test
Connect the pulse generator to a power splitter.
Connect one side of the power splitter to Input A on the
counter using a coaxial cable.
Connect the other side of the power splitter to Input B
on the counter.
Settings for the pulse generator:
Amplitude = 2 V
PP
, (high level +2V and low level 0 V)
Period = approx. 1 ms
Duration = approx. 50 ns
Rise time = 2 ns
Restore the timer/counter's default settings and make the fol
-
lowing changes:
Function =
Time A-B
Press STAT/PLOT key to the right of the display.
Settings for
INPUT A and INPUT B:
50 W input impedance
MANual trigger level
Performance Check, Short Form Specification Test 2-5
Settings V
max
V
min
Pass/Fail
Input A Input B
INPUT A, DC, 50 W
+950 mV
+50 mV
AC
+450 mV
-450 mV
10X
+0.45 V
-0.45 V
1MW
>+0.45 V
<-0.45 V
Table 2-6 Input controls check.
Oscillator Frequency Readout Suitable
Reference
P/F
Standard
(PM6690/_1_)
10.00000000 MHz ± 150 Hz
908
OCXO
(PM6690/_5_)
10.00000000 MHz ± 1Hz
909
OCXO
(PM6690/_6_)
10.00000000 MHz ± 0.25 Hz
909
Table 2-7 Acceptance test for oscillators.

Table of Contents

Other manuals for Fluke PM6690

Related product manuals