Pull the interior unit out of the cover.
Reinstalling the Cover
Push the counter gently back into the cover.
Turn it upside down
Fit the two screws at the bottom.
Fit the two rear feet with their screws to the rear panel.
Fit the four decorative plugs.
Fan
Disconnect the power cable.
Remove the cover from the counter. See page 3-2.
Remove the four screws, nuts and washers that fix the
fan to the rear panel.
Disconnect the fan power supply connector from the
main PCB and remove the fan.
When reinstalling the fan, make sure the air-flow arrow
on the fan points to the rear of the counter and the black
wire is oriented toward the rear panel.
Prescaler Options
Disconnect the power cable.
Remove the cover from the counter. See page 3-2.
Disconnect the cable from the mini-coax connector on
the RF input.
Press the clips apart and lift the RF input PCA straight
up and out.
When installing the RF input, make sure the connector
pins fit exactly in the holes of the connector housing.
Disassembly, Fan 3-3
Figure 3-4 The fan is fixed with four screws and nuts. The 3 GHz & 8 GHz prescalers are fixed with snap-in clips. The OCXO is sol
-
dered to the main board.
4 screws to remove
C
Figure 3-5 Remove the screws and pull out the main
PCB.