EasyManua.ls Logo

Fluke PM6690 - Sensitivity, Power Supply, and External Reference Adjustments

Fluke PM6690
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
n
Adjustment
Connect the probe tip to the center hole of J7.The sur
-
rounding holes are suitable ground connections. See Fig
-
ure 7-2.
Adjust R196 until the pulses on the oscilloscope look
the same. See Figure 7-3. Use a screwdriver type of ad
-
justment tool.
Check Step Response X1, repeat Step Response X1 and
Linearity several times if necessary.
Repeat all steps above for Input B on the tim er/counter. Con
-
nect the probe tip to J9. Adjust R268.
Sensitivity
n
Setup
Keep the settings on the counter from Linearity.
n
Adjustment
Connect the DMM to the test points X34 (+) and X35
(-). See Figure 7-2.
Adjust R197 until the DMM reads +17 ± 0.2 mV. Use a
screwdriver type of adjustment tool.
Don't forget to resolder the RF shield.
Power Supply
n
Setup
Do as described under Preparations above.
Select default settings by keying in
USER OPT ®
Save/Recall ® Recall Setup ® Default
n
Adjustment
Connect the DMM to test point X9 and ground (metal
shield lid over input amplifiers). See Figure 6-4 and
Figure 7-4.
Adjust the +5 V trimmer on the AC/DC power supply
board with a screwdriver type of adjustment tool until
the DMM reads +5.10 ± 0.005 V.
Check that the voltage from the AC/DC power supply at
X26 is +15 ± 0.7 V.
Check that the voltage from the AC/DC power supply at
X21 is -15 ± 0.7 V.
Note: Adjusting the +5 V does not affect other settings in the
instrument.
External Reference Input
n
Setup
Do as described under Preparations above.
Select default settings by keying in
USER OPT ®
Save/Recall ® Recall Setup ® Default
Connect a 1 MHz sinewave signal with amplitude 0.1
V
rms
(0.28 V
pp
) into 50 W from an LF synthesizer to the
external reference input. Use a 50 W termination on the
input.
Set an oscilloscope with X10 probe to 10 ms/div,
100 mV/div (including probe) and AC coupling.
n
Adjustment
Connect the probe to X19. There are two ground pads in
the vicinity. See Figure 7-5.
Adjust C442 to maximum amplitude. Use a screwdriver
type of adjustment tool.
Power Supply 7-13
+5 V adj.
J17
X26
X9
X21
GND
AC/DC BOARD
RF SHIELD
Figure 7-4 Adjusting +5 V and checking +15 V and -15 V.
C442
X19
GND
GND
AC/DC BOARD
J22
J25
J23
J26
J21
Figure 7-5 Adjusting the external reference input.

Table of Contents

Other manuals for Fluke PM6690

Related product manuals