EasyManua.ls Logo

Fluke PM6690 - HOLD OFF Function and RF Options Check

Fluke PM6690
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Check that the timer/counter performs the correct mea
-
surement by displaying the result as shown under the
“Display” column in Table 2-8.
Select function via
MEAS FUNC
Check of HOLD OFF
Function
Recall the DEFAULT settings.
Select
Period Single A.
Select the following common timer/counter settings for both
Input A and Input B via the hard menu keys
INPUT A and IN
-
PUT B
:
50 W impedance.
DC coupling.
MANual trigger, x1 attenuation.
Press
SETTINGS and activate Hold Off. Select Hold Off
On
and set the Trigger Hold Off time to the minimum
value 20 ns.
Connect the rear panel output marked 10 MHz OUT to
Input A.
Increase the
Trigger Hold Off time in steps by means of
the
UP cursor key and note the results:
If Trigger Hold Off ti me <100 ns the result is about 100 ns,
i.e. the same value as without Hold Off.
If Trigger Hold Off time >100 ns the result is about 100 ns +
Trig Hold Off time.
Connect the signal to Input B, select
Period Single B and repeat the tests above.
RF Options
Input C Check
To verify the specification of the different RF prescalers
(Input C), use the following basic test setup:
Connect the output of a signal generator covering the
specified frequency range to the RF input of the counter.
Connect the 10 MHz REF OUT from the generator to
the EXT REF IN on the rear panel of the counter.
Choose
Meas Ref from the SETTINGS menu and select
External.
Choose
Freq C from the MEAS FUNC menu.
Generate a sine wave in accordance with the tables.
Verify that the counter is counting correctly. (The last
digits will be unstable)
Performance Check, Check of HOLD OFF Function 2-7
Frequency Amplitude P/F
MHz mV
RMS
dBm
100-300 20 -21
300-2500 10 -27
2500-2700 20 -21
2700-3000 40 -15
Table 2-9 RF input sensitivity,
3 GHz Option.
Frequency Amplitude P/F
MHz mV
RMS
dBm
300-500 20 -21
500-3000 10 -27
3000-4500 20 -21
4500-6000 40 -15
6000-8000 80 -9
Table 2-10 RF input sensitivity,
8 GHz Option.

Table of Contents

Other manuals for Fluke PM6690

Related product manuals