5:2 User Manual 6008 1286 / Rev. 8
5.4 Replacement of Lamp
The instrument should be well cleaned before replacement of the lamp to prevent
dust from damaging the reflector. The replacement should be performed in a clean
environment.
Warning
The lamp module can be very warm if the operating temperature
is high. Cotton gloves must be used when replacing the lamp to
avoid burn injuries.
Caution
Do not touch the lamp glass or reflector or let any rough surface
come into contact with the lamp glass. A microscopic scratch in
the glass might cause a lamp explosion later.
Caution
The lamp can be damaged by fingerprints and oily residues.
Cotton gloves must be used when replacing the lamp to avoid
any damages.
Replace the halogen lamp as follows:
1. Switch off the instrument. Disconnect from mains.
2. Unscrew the two screws 1 fastening the cover plate using the supplied 3 mm
allen key.
Fig. 5:1Cover plate to lamp assembly