NIRS™ DA1650
User Manual 6008 1286 / Rev. 8 8:1
8 Technical Specifications
8.1 Technical Data
8.2 Installation Requirements
NIRS™ DA1650
Dimensions (W x D x H) 230 x 530 x 280 mm
Weight 16 kg
Fuse 6.3 A
Power consumption 70 VA
Noise level < 70 dB(A)
Degree of protection IP 65
NIRS™ DA1650
Voltage supply
100-240 V AC *), frequency 50-60 Hz, Class 1,
protective earth
Ambient temperature 5-40 °C
Storage temperature -20 °C to 70 °C
Ambient humidity < 93 % RH
Mechanical environment Stationary and moveable occasionally
EMC environment Laboratory use, Industry requirements
Operation Indoor use
Altitude Up to 2000 m
Transient over-voltage According to category II
Pollution Degree 2
Vibrations, random
(two different spectras)
0.19 G
rms
at 10-150 Hz to IEC 60068-2-64 spectra
0.19 G
rms
at 10-1250 Hz to FOSS internal spectra
(more information available on request)
*) Mains supply voltage fluctuations not exceeding ±10 % of the rated voltage.