Overall Accuracy against YSI Reference
Accuracy was measured by comparing the absolute relative dierence between the System and
reference YSI glucose values. The absolute relative dierence measures the level of disagreement
between the System and the reference value, but does not tell you whether the System glucose
value was, on average, higher or lower than the reference glucose value. The Mean Absolute
Relative Dierence gives an indication of the average percent disagreement between the GM and
the reference. Table 4 shows the absolute dierence measures by glucose level. Overall the Mean
Absolute Relative Dierence was 12.3% for the comparison with YSI reference. The Median Absolute
Relative Dierence shows that half of the time the System was within 10.1% of the YSI reference.
Table 4: Dierence Measures by YSI Reference Glucose Levels
YSI Reference
Reference Glucose
Level (mg/dL)
Number of GM-
Reference Pairs
Median Absolute
Relative Dierence (%)
Mean Absolute Relative
Dierence (%)
Overall 12323 10.1 12.3
40-50* 30 14.2 15.8
51-80* 505 12.5 15.5
81-180 7373 10.7 12.9
181-300 4115 8.7 10.1
301-400 286 7.8 8.8
401-500 14 4.2 7.2
* For reference values ≤ 80 mg/dL, the mean and median absolute dierences (mg/dL) are presented instead of
mean and median absolute relative dierences (%).
48 49
ART26944_rev-C_manual.indd 48-49 9/23/16 9:19 AM