Chapter 6: Maintenance
Fresenius Kabi CompoSeal Universal OP-EN 6/07.13 15
This will disengage the lever.
Put down the hand sealer. The lever must be
in the position shown in the illustration.
Push the sealing head upwards with your
thumb (1), while exerting a counter-pressure
on the immovable part with your index finger
(2), slightly tilt and remove the sealing head
by pulling it to the front (3).
Remove the leaf spring (1) by pulling it to the
front.
Remove the lever.
1
2
2
3
1