EasyManua.ls Logo

FT Technologies acures FT722 - Page 46

FT Technologies acures FT722
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 4 Service, Configuration & Testing
46 FT722 & FT742 (RS485) FF & PM Sensors User Manual
UCT (User Calibration Table) Tab
The UCT window provides the user with a simple method of changing the calibration performance of the sensor.
Refer to the UCT commands detailed in the user manual for further information.
Acu-Vis 2.0 will download any existing UCT settings and display them on the table and graph.
The easiest method of changing the UCT is to use the ‘Standard % Offset” buttons to establish a calibration
change.
When the calibration change has been selected press the orange ‘Save UCT to Sensor’ button and ensure the
‘UCT Status’ is set to Enabled.
Caution: This feature will change the calibration profile of the sensor and may compromise survey data
and/or system performance
Figure 43: UCT Modification Window

Table of Contents

Related product manuals