27
INZ-TN2ZRE-E
 
This mode is used to set calibration concentration and actions. The calibration setting involves cali-
bration concentration, zero calibration, calibra-
tion range and auto calibration component/range.
In the “Calibration Parameters” screen that ap-
pears, the data shown at right is illustrated.

It allows you to set concentrations of the stan-
dard gas (zero and span) of each Ch used for
calibration.
(1) Select < User mode > → < Calibration
parameters > → < Calibration value >.
“Caribration Value Settings” screen ap-
pears as shown at right.
(2) Select the Ch you want to change by
pressing the
or the key.
Press the
ENT
key and cursor moves preceding the
value.
(3) Select the concentration item you want
to make the setting by pressing the
,
key or the key (movable within
the selected Ch).
Then press the
ENT
key, and the selected
value is highlighted.
( )
ENT
( )
ENT