EasyManua.ls Logo

Fuji Electric ZRE-2 - Page 40

Fuji Electric ZRE-2
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
INZ-TN2ZRE-E
Press the key to exit automatic calibration component/range setting, and the previous
screen appears.
To close the setting
ESC
Operation by setting
Auto calibration is performed under the following rules.
1. Zero calibration is performed at the same time, for the Ch (component) with which “enable”
is selected at the time of auto calibration and auto zero calibration.
2. Span calibration is performed in the order from smallest Ch No., for the Ch (component)
with which “enable” is selected at the time of auto calibration.
ZERO calibration on auto calibration and auto zero calibration of the component with
which “enable” is selected are performed in batch irrespective of the description in
“6.2.2 Setting of manual zero calibration.”
Note

Table of Contents

Related product manuals