▪ Current: 1VDC per 10uA
▪ Current: 1VDC per 100uA
▪ Current: 1VDC per 1mA
▪ Current: 1VDC per 10mA
▪ MidiVac Emulation
The logarithmic output calculates the log base 10 of the selected
measurement (current or pressure) and adds an oset in order to make the
voltage output positive. (The Log base 10 of a number less than 1 is
negative.) When logarithmic output is selected, an oset must also be
specied. The higher the oset value, the higher the output.
Figure 14. System output supply
6.4. Supply menu (supply)
Figure 15. Supply menu
Arc Detection (Arc)
Arc detection provides protection for the ion pump by lowering high voltage
during arcing. The default setting is o.
Figure 16. Arc processing
Power Loss Recovery (Auto HV)
The software contains a power loss recovery feature to restart the system if
the system was turned o due to a power failure.
▪ Disabled (default option): If input power is disrupted, the controller will
not power up when input power is restored.
04/2021 - ©Gamma
Page 21900026_F
900026_F - Display menu