Display Message Display Meaning
31: HV Control Error Supply HV control mechanism malfunctioning. On/O state
is malfunctioning.
32: Current Control Err Supply current control mechanism malfunctioning.
33: Current Measure Err Supply current measuring mechanism malfunctioning.
34: Voltage Control Err Supply HV control mechanism malfunctioning. Voltage
output level is malfunctioning.
35: Voltage Measure Err Supply voltage measuring mechanism malfunctioning.
37: HV Not Installed Internal boards polarity mismatch.
37: Input Voltage Error HV module missing.
38: Input Voltage Error Input power voltage outside 22-26 V d.c. range. HV
operation disabled. When running in battery mode, this
restriction is o.