Calibration
34
4) Open Gamry Framework™ software. Select Experiment > Named Script...
The Select a script to run window appears.
5) From the list of scripts, choose calcable.exp,
then click the Open button.
The Cable Capacitance Calibration window
appears.
6) In the Cable Tag field, enter a
unique name for the cable you
are calibrating.
Choose the desired Action radio
button:
• To calibrate the cable, choose
Cal Cable.
• To reset the vales to zero (if,
say, the calibration doesn’t work),
choose Zero Values.
Click the OK button.
The Performance Tips window appears: