Page 14 of 53 Copyright © 2007, General Electric Company 2023486-201 Rev C
BATTERY INSTALLATION
1. With the label on the battery facing the AED Pro battery compartment, insert the battery as
shown in the drawing.
2. Push the latched end of the battery firmly into the AED Pro, as shown in the drawing, until the
battery snaps into place. The exposed side of the battery should be flush with the outside of the
AED Pro case.
3. Open the lid for 5 seconds to initiate self-test. If the battery is installed properly, the STATUS
INDICATOR will turn GREEN. Close the lid.
Note: Batteries with part number 2023489-001 (9144) and 2023681-001 (9145) are for use only with
Responder AED Pro and should not be used with other AEDs.