EasyManua.ls Logo

GE AED Pro - RHYTHMX AED ECG Analysis Algorithm

GE AED Pro
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 7: Accessories
2023486-201 Rev C © 2007 General Electric Company Page 43 of 53
RHYTHMX
®
AED ECG ANALYSIS ALGORITHM
The RHYTHMx AED ECG analysis algorithm provides superior ECG detection capabilities, allowing it to be placed on
patients at risk for sudden cardiac arrest. The features available with the AED Pro include the following:
Detection Rate
Asystole Threshold
Noise Detection
Non-Committed Shock
Synchronized Shock
Pacemaker Pulse Rejection
SVT Discriminators
Supraventricular Tachycardia (SVT) Rate
DETECTION RATE
All ventricular fibrillation (VF) and ventricular tachycardia (VT) rhythms at or above this rate will be classified as shockable.
All rhythms below this rate will be classified as non-shockable. This rate is configurable between 120 bpm (beats per
minute) and 240 bpm. Service can change this rate using the ServiceLink software. The default Detection Rate is 160
bpm. The Responder AED Pro detection rate is 160 bpm.
ASYSTOLE THRESHOLD
The Asystole baseline-to-peak threshold is set at 0.08 mV. ECG rhythms at or below 0.08 mV will be classified as
Asystole and will not be shockable.
NOISE DETECTION
The AED Pro will detect noise artifact in the ECG. Noise could be introduced by excessive moving of the patient or
electronic noise from external sources like cellular and radiotelephones. When noise is detected, the AED Pro will issue
the prompt “ANALYSIS INTERRUPTED. STOP PATIENT MOTION” to warn the operator. The AED Pro will then proceed
to reanalyze the rhythm and continue with the rescue.
NON-COMMITTED SHOCK
After the AED Pro advises a shock, it continues to monitor the patient ECG rhythm. If the patient’s rhythm changes to a
non-shockable rhythm before the actual shock is delivered, the AED Pro will advise that the rhythm has changed and
issue the prompt “RHYTHM CHANGED. SHOCK CANCELLED.” The AED Pro will enter the CPR mode and prompt,
"START CPR".
SYNCHRONIZED SHOCK
The AED Pro is designed to synchronize shock delivery on the R-wave. The AED Pro will automatically attempt to
synchronize to the R-wave. If delivery cannot be synchronized within one second, a non-synchronized shock will be
delivered.
PACEMAKER PULSE DETECTION
The AED Pro contains pacemaker pulse detection circuitry to detect pulses from an implanted pacemaker.

Other manuals for GE AED Pro

Related product manuals