Page 42 of 53 Copyright © 2007, General Electric Company 2023486-201 Rev C
BATTERIES AND CAPACITOR CHARGE TIMES
A new battery typically takes 10 seconds to charge the AED Pro to maximum energy.
A battery with reduced capacity causes the red LED light to initially turn ON and typically takes 13 seconds to charge a
fully discharged AED Pro to maximum energy.
The maximum time from “Power On” to “Ready to Shock” is 28 seconds for a new rescue.
The maximum time from “Analyze” to “Ready to Shock” is 22 seconds for a new rescue.
AED Pro SELF-TEST SEQUENCE
Frequency of Self-Test What is Tested?
Daily
Weekly
Monthly (every 28 days)
Open Lid (when lid is opened)
Close Lid (when lid is closed)
Battery, pads, internal electronics, SHOCK button, and software (no charge).
Battery, pads, internal electronics, SHOCK button, and software (partial charge).
Battery under load, pads, internal electronics, full-energy charge cycle, SHOCK
button, and software (full charge).
Battery, pads, internal electronics, SHOCK button, and software.
Battery, pads, internal electronics, SHOCK button, and software.