Section 7: Accessories
2023486-201 Rev C © 2007 General Electric Company Page 41 of 53
PADS (ADULT)
Self-adhesive, disposable defibrillation pads
Minimum combined surface area: 228cm
2
Extended length of lead wire: 1.3m
LITHIUM SULFUR DIOXIDE BATTERY SPECIFICATIONS
Output voltage: 12VDC (max)
Batteries are non-rechargeable
Lithium contents: 9.2g (max)
Check local regulations for disposal information
Model Estimated Shelf Life Warranty Typical Shocks
2023681 Lithium
Sulfur Dioxide
5 Years
1 Year of 12 hours
of use, whichever
occurs first
Up to 290 shocks
The battery operating life depends on the type of battery, actual usage and environmental factors.
RECHARGEABLE BATTERY SPECIFICATIONS
Battery Voltage: 11.1V
Chemistry: Lithium-ion. Refer to local regulations.
Compatibility: Responder AED Pro Model 2023440
Battery Capacity: 60 shocks minimum (100 shocks typical) or 3 hours minimum (6 hours typical) of ECG display
time.
Battery Charge Time: 3 hours for stated capacity, 4.5 hours to fully charge completely depleted battery.
Battery Standby: 6 months
Battery Life: 2.5 years or 300 Battery charge-discharge cycles, whichever comes first.
Battery Weight: .52kg (1 lb. 3 oz)
BATTERY CHARGER
Power Requirements: 90 to 132 VAC or 198 to 264 VAC at 47 to 63 Hz
The Charger operates from, and accepts standard IEC mains power cables.