Page 18 of 53 Copyright © 2007, General Electric Company 2023486-201 Rev C
B. NUMBER OF SHOCKS DELIVERED Indicator
This indicator counts and displays the number of shocks delivered.
C. ELAPSED RESCUE TIME Indicator
This indicator times and displays the elapsed rescue time.
Note: There is a 3 second delay between the time the AED Pro lid is opened and the start of the
rescue (when the lid was first opened).
D. HEART RATE Indicator
This indicator displays the patient’s heart rate.
E. ECG Display
Four and a half seconds of the patient’s ECG is displayed.
F. PAD PLACEMENT Display
Visually assists the operator with pad placement with the directions for use. Appropriate text
prompts are also displayed.
G. TEXT Display
The text display has 2 lines of text. It provides the operator with information regarding system
initialization, text version of the voice prompts and data during a rescue, and diagnostics.
System initialization occurs when the lid is first opened. The text display shows the operator the
identifiers for the internal code, voice prompts and text prompts versions.
H. CPR Counter
During CPR, a countdown timer will be displayed.
I. SERVICE Indicator
When displayed, indicates that service is required that can only be performed by qualified service
personnel.
Note: When Status Indicator is RED or Service Indicator is illuminated, device cannot be
used to perform a rescue.
J. PAD Indicator
The pad indicator will flash with a voice and text prompt indication “Check Pads” when one of the following
occurs:
• Pads are not properly connected to the AED Pro
• Pads are not within operational specifications (cold, dried, damaged)
• Pads are not connected properly to the patient during the rescue.
K. Z-BAR Indicator
The Z-Bar provides a relative visual graphical indicator of the total transthoracic impedance between the two
defibrillation pads. The Z-Bar is used in the assessment of:
• Adequate pad placement
• Pad quality and integrity
• Pad adhesion to the patient’s skin
• Proper pad connection to the AED Pro
• Provides for quick assessment between pad off and pads shorted