Page 22 of 53 Copyright © 2007, General Electric Company 2023486-201 Rev C
ADVANCED FEATURE PROMPTS
VOICE PROMPT
TEXT DISPLAY SITUATION
“Entering Manual Mode. Press Button
Again to Confirm”
“Manual Mode. Charging”
“Manual Mode Not
Confirmed.”
“If Rhythm is Shockable, Press
SHOCK Button to Deliver Therapy.”
“Shockable Rhythm. Attach
Defibrillation Pads.”
“Device Will Disarm in :30”
MANUAL MODE
PRESS BUTTON TO CONFIRM
MANUAL MODE
CHARGING
MANUAL MODE
NOT CONFIRMED
IF SHOCKABLE RHYTHM
PRESS SHOCK BUTTON
SHOCKABLE RHYTHM
ATTACH DEFIBRILLATION PADS
DEVICE WILL DISARM IN :30
Prompt issued after ALS presses the
MANUAL button once to initiate the
manual mode.
Prompt issued after ALS presses the
MANUAL button again to confirm.
Prompt issued when the MANUAL
button is not pressed a second time
within five seconds, the device stays
in AED Pro mode.
Prompt issued when in manual mode,
prompts ALS personnel to press
SHOCK button if ECG indicates a
shockable rhythm.
Prompt issued when the device is
performing ongoing ECG monitoring
via the ECG Patient Cable Kit and
detects a shockable rhythm.
Should the rescuer go into manual
mode and decide that AED mode is
more appropriate, the AED Pro will
revert back to AED mode 30 seconds
after charging is complete. The
seconds will count down from 30
on the display.
When "Remain in manual mode" has
been enabled (Using ServiceLink
software). The AED will disarm but
remain in Manual Mode. See page 27