5.3 Select tray
The following table describes the suitable tray and accessory for each capturing method.
Use the correct tray to the selected exposure for optimal image quality.
Note:
Available in
configura-
tion
Tray and accessorySample typeImage capturing method
680Black Tray
•
Membrane
Chemiluminescence
680 UV
680 QC
680 RGB
680Black Tray +
•
Membrane
Chemiluminescence
with colorimetric marker 680 UVWhite Insert
680 QC
680 RGB
680Black Tray +
•
Membrane
•
Gel
Colorimetric
Epi illumination 680 UV
680 QC
White Insert or,
White Trans Tray
680 RGB
680 QCWhite Trans Tray +
•
Gel
Colorimetric
Trans-illumination 680 RGBDiffuser Board
680 RGBBlack Tray
•
Gel
•
Membrane
Fluorescence RGB
Epi-illumination
680 UVUV Trans Tray
(transparent) +
•
Gel
Fluorescence
Trans-UV 680 QC
Gel sheet
680 RGB