4 - 6 Maintenance
K0260 Issue No. 8
Hand Terminal Cable Tests
Carry out the following check as detailed in Testing and Fault Finding, Section 5.
Measure continuity using the DVM (item 1, Table 4-3) set to an appropriate range.
Measure the continuity between corresponding pins at each end of each cable assembly.
The measured resistance must not exceed 0.10 .
Measure insulation using the Insulation tester (item 2, Table 4-3), set to an appropriate
range.
Measure the insulation resistance between the shell of the connector and all individual pins
in turn.
Measure the insulation resistance between the shell of the appropriate connectors and all
pins in turn.
Measure also the resistance between individual pairs of pins (i.e., 1-2, 1-3, 1-4, 1-5, 1-6, 2-
3, 2-4, 2-5, 2-6, 3-4, 3-5, 3-6, 4-5, 4-6 and 5-6). In all cases the resistance must exceed 10
Mohms.