Druck ADTS 505 User Manual 5 - 15
K0260 Issue No. 8
5
TABLE 5-3 ERROR MESSAGES (CONTINUED)
No. Message Probable Cause Action
219 Abort sequence started by user
Abort key pressed on
hand terminal
Wait until system completes abort
sequence
220
"Cannot Modify Altitude
Correction in Pressure Units"
Altitude Correction
Change units in setup to aero units
221 "Changed Aim to ARINC Limit"
Value out of ARINC limit
range
Re-enter value within allowed range
222 Calibration data failed
Incorrect pressure
value entered during
calibration
Check and enter correct pressure
value for reference pressure point
223 "Cannot enter EPR in Aero Units"
EPR selected with aero
units displayed
Change units in setup to pressure
units then select EPR
227 "Changed Aim to Current Limit" Value out of limit range
Re-enter value within allowed range
228 "Limit Overlap"
Value conflicts with
limit ranges
Re-enter value within allowed range
229 "Limit Saved"
Value out of acceptable
range
Re-enter value within allowed range
241 "Ground Request in Leak Mode"
System must be in
control mode
Select control mode
243
"LDK has an invalid EEPROM
archive"
Invalid/corrupted data
- recovery in progress
Check custom limit sets not lost
245
"Controller has an invalid EEPROM
archive"
Invalid/corrupted data
- recovery in progress
Check/correct calibration
247
"Handset has an invalid EEPROM
archive"
Invalid/corrupted data
- recovery in progress
250
"Vacuum pump processor has
failed"
Internal fault
Return unit to repair depot
251
"Vacuum pump lead is
disconnected"
-
Return unit to repair depot
252
"Vacuum pump transistor is faulty
or drive inhibited"
Internal fault
Return unit to repair depot